|
|
The ILX Lightwave FPM-8210H Fiber Optic Power Meter has a single input port for both connectorized and bare fiber measurements. Virtually insensitive to polarization state or to pattern changes from fiber orientation, these meters deliver repeatable high resolution results. ILX Lightwave engineered the FPM-8210H for the production test workstation. The innovative fiber optic head design makes reliable fiber endface positioning a routine task, whether or not the fiber has a connector. With the CA-120 Bare Fiber Adapter, this meter is compatible with either ILX Lightwave or Agilent bare fiber holders. Specifications.
Wavelength Range: 800 to 1650 nm. Power Range: -50 to +30 dBm. Damage Threshold: >+40 dBm. Accuracy Reference Conditions: ±2.5%. Accuracy Operating Conditions: ±5.0%. Polarization Dependent Response: ±0.002 dB, typical. Measurement Repeatability: ±0.003 dB, typical. Sensor Type: InGaAs. Noise: <=500 pW p-p (1200 at 1630nm) typical <250pW p-p at 1310 and 1550 nm. Temperature Coefficient: Typical ±0.2%/°C. Linearity: ±0.04 dB, ±500 pW (-40dBm to 30dBm). Analog Output Bandwidth: Typical 0 to 10 Hz typical, Voltage: 0 to 10V, Impedance: Typical 1000 ohm.
|